Jerome® J605
Hydrogen Sulfide Analyzer
This next-generation of the renowned Jerome® family of vapor analyzers maintains all the features, reliability
and simplicity that have established Jerome® as an industry leader in the hydrogen sulfide and mercury vapor
gas detection markets for more than 20 years. Significant performance enhancements and new features now provide a new
alternative for industrial hygienists and consultants to detect, measure and report levels of hydrogen sulfide vapor for
environmental, industrial and waste water treatment measurements.
The J605 has an accuracy of ±1 ppb at the calibration of 5 ppb - fullfilling the needs of the tightest of EPA requirements
for hydrogen sulfide testing. The enhanced J605 has an incredible resolution of 20ppt and detects hydrogen sulfide vapor
levels as low as 3 ppb with confidence. The J605 is housed in a lighter and more ergonomically designed case, setting a new
precedence for low-level, portable monitors. The industry-proven, gold film sensor has been upgraded and improved, resulting in
longer life expectancies. The larger display, onboard data logging, auto zeroing, and battery-powered sensor regeneration will
make it easier to use and maintain the instrument in the field. Communication options and data acquisition have become more
simplified with USB communication, SCADA compatibility via 4-20mA output, and menu-driven programming options.
- Leak Detection
- Quality Control
- Ambient Air Analysis
- Accuracy Check for other Hydrogen Sulfide Monitors and Control Systems
- Applied Research
- Scrubber Efficiency Testing
- Hydrogen Sulfide Source Detection
- Control Room Corrosion Monitoring
- Landfills
- Agricultural Applications
- Regulatory Compliance and Permitting
- Rugged and easy to operate
- Battery powered sensor regeneration (requires external battery pack)
- Auto date/time stamp
- On board data logging 20,000 samples
- Inherently stable gold film sensor
- Pressure sensitive membrane switch operation
- Rechargeable internal battery pack for portability
- Wide detection range allows for multiple applications
- Automatic LCD backlight during low light conditions
- Survey mode for rapid source detection of hydrogen sulfide concentrations
- Microprocessor ensures a linear response throughout the entire range of the sensor
- 4-20 mA output for SCADA systems
- No PC Software Required
- 20,000 Data Points
- Stores Date, Time and Locations
- On-Board Data Logging
- Optional USB interface for data transfer and printing capabilities
- 18 Hour Battery Life
- Detects H2S levels as low as 3 ppb
- Fulfills EPA requirements
| Resolution | 20 ppt in Range 0 |
| Detection Range | 3 ppb -10 ppm |
| Accuracy/Precision | Range 0: ±1 ppb at 5 ppb/10%RSD Range 0: ±3 ppb at 50 ppb/5%RSD
Range 1: ±0.03 ppm at 0.5 ppm /5%RSD Range 2: ±0.3 ppm at 5.0 ppm/5%RSD |
| Response Time |
| Manual Range Selection | |
| Range 2 (1.0 to 10.0 ppm) | 12 seconds |
| Range 1 (0.10 to 1.0 ppm) | 18 seconds |
| Range 0 (3 to 100 ppb) | 12 to 27 seconds |
| Auto Range Mode | 12 to 52 seconds |
|
| Flow Rate | 150 ml/min or 0.15 l/min |
| Power Requirements | 12V DC/power adapter |
| Internal Battery Pack | Rechargeable nickel metal hydride (NiMH) |
| Environmental Range | 0-40°C, noncondensing, nonexplosive |
| Interface | USB |
| Dimensions | 11" W x 6" L x 6.5" H/ 28 cm L x 16 cm W x 17 cm H |
| Weight | 5.4 lbs / 2.5 kg |
| Warranty | 1 year, factory parts and labor |
| Certifications | European Communities (CE) and TUV: UL 61010-1:2004 standard and CSA-C22.2 NO. 61010-1-04 |
- Functional Test Module for instrument verification in the field
- Maintenance Kit for routine maintenance and upkeep
- Hard or Soft Field Carrying Cases for versatile handling and additional storage
- Car Adapter, External Battery Pack