The Jerome®
471 mercury vapor analyzer is the most recent addition to Arizona Instrument's Jerome family of vapor detection
instruments. With over 20 years experience and unparalleled recognition in the portable vapor analysis market, Arizona
Instrument LLC is introducing the next generation Jerome®.
Capable of detecting accurately down to 0.030 µg/m3, the 471 utilizes the UV Atomic Absorption
spectrometry method of mercury vapor assessment. Accuracies are maintained by means of an
internal reference cell. Interferences common in other technologies are greatly reduced due to the
absorption specificity of the mercury atom.
Portable and easy-to-use, the 471 will provide up to 16 hours of field testing on one battery charge.
The on-board controller will allow the user to easily navigate the touch screen set-up and testing
modes. The user-programmable Threshold Settings feature allows the operator to select the limits
at which testing will be conducted. A padded, adjustable shoulder strap and detachable probe aid in
providing comfort and minimizing fatigue for the user during long periods of testing.
| 471 Features |
- Ultra-low Mercury detection level in preparation for more demanding EPA regulations
- Wide detection range for applications ranging from emergency spill response to post clean-up
- Compliance level or Scientific level user modes
- Programmable Industrial or Residential threshold limit settings
- 5-point Factory Calibration; 0.2, 0.5, 2, 10 and 25 µg/m3
- User-friendly software for ease-of-use and data reporting
- Touch screen operation
- Linear response throughout entire range
- Internal data logging with available USB connection for data transfer to a laptop or PC
- Auto zeroing
- Built-in user’s Help guide
|
| Optional Equipment |
- Functional Test Kit
- Maintenance Kit for routine maintenance
|
| Applications |
- Mercury Spill Response & Cleanup
- Mercury Exclusion Testing
|
- Hazardous Waste Sites
- Fluorescent Lamps Disposal & Recycling
|
| Specifications |
| Resolution |
0.001 µg/m3 |
| Detection Range |
0.030 µg/m3 to 250 µg/m3 |
| Precision |
5% Relative Standard Deviation at 2 µg/m3
20% RSD at 0.2 µg/m3 (5 Samples) |
| Accuracy |
±2% at 25 µg/m3
±20% at 0.2 µg/m3 |
| Data Storage Capacity |
100,000 samples |
| Power |
Internal battery for up to 16 hours of portable operation.
External battery charger (100-240 V) |
| Environmental Range |
0 to +40°C non-condensing, non-explosive |
| Interface |
USB (for downloading data to a PC) |
| Dimensions |
6" W x 19" L x 8" H /152 mm W x 483 mm L x 203 mm H |
| Weight |
17.5 lb / 7.9 kg |
| Warranty |
1 year, factory parts and labor |
24 Hour Customer Support
|
|