COMPUTRAC® Vapor Pro RX™
Moisture Analyzer


Designed Exclusively for Lyophilized Pharmaceuticals!

No toxic reagents or specialized glassware
Maintains the sealed vial integrity
Automatic, menu-driven operation requires minimal user skills
Automatic sample loading
Accepts multiple bottle sizes
Low detection range, down to 100 ppm
Wide temperature range, programmable start and test temperatures from 25- 275 °C
Moisture specific sensor, eliminates volatile interferences associated with LOD technology
Self-diagnostics alert operator of any instrument problems
Security feature restricts access to program and instrument settings
Optional multi-point temperature calibration module utilizes NIST-traceable RTD to ± 1 °C
Optional validation package in Adobe Acrobat format and validation services available

Vapor Pro RX Specifications:
Sample Size Dependent upon content of vial
Vial Size Range 13-37 mm diameter, 29.5-65 mm long
Resolution 1 ppm
Moisture Range 100 ppm or 0.01% to 100% (dependent upon sample size)
Results Display % Moisture, ppm Moisture, or µg Water
Ending Criteria Predict, Rate, and Time
Heating Range 25 - 275 °C, Set in 1 °C increments and maintained to ± 1 °C
Test Parameter Memory 102 Programs
Program Identification Alphanumeric and/or symbols; programmable prompts for lot or product ID numbers
Test Result Memory Data from last 30 tests
Statistical Calculation Mean, Standard Deviation, and Coefficient of Variation
Power Requirements 100-120V ~ 50/60 Hz, 1 amp (standby) or 8 amps (heat on),
or 220-240V ~ 50/60 Hz, 0.5 amp (standby) or 4 amps (heat on)
Interfaces RS-232 (9-pin standard) for balance or PC interface; parallel printer (LPT - PC
standard); 5 pin DIN connector for 101 key PC keyboard
Dimensions 146mm H x 380mm W x 370mm D; 5.75" H x 15" W x 14.5" D
Weight 8.6 kg; 19 pounds
Warranty Two years, factory parts and labor (one year international)
Certifications U.S. and Canadian Underwriters Laboratory; European Communities - CE
Accessories Keyboard; Printer; Analytical Balance; Carrying Case; Compressed Air Pump; Multi-Point Temperature Calibration Module; Validation Package

Additional Features:
Easy setup of product methods to ensure optimum test results
Programmable starting temperatures and maintained within ± 1 °C
Continuous display of test time, current temperature, programmed test temperature, current microgram water measurement, and calculated moisture content
Flexible ending criteria to ensure accurate test results
Statistical features allow calculation of mean, standard deviation and coefficient of variation
Real-time graph of moisture curve and rate of moisture loss

 

Trial Instruments Available • Free Application Development
24 Hour Customer Support • Free Loaner Service